LCP 酸鹼度感測器 Liquid Crystal Polymer pH Sensors
產品編號
Hach pH感測器搭配Hach sc200 或 sc1000數位水質控制平台協助您線上監測酸鹼度

LCP (Liquid Crystal Polymer), Convertible Mounting Style, 4.5m Analog Cable, 5-Wire with Built-In Preamplifier, General Purpose Glass pH Electrode

Exceptional Performance with the Differential Electrode Measurement Technique
Lower Maintenance Needs with the Replaceable Salt Bridge

Built-in Preamp or Two-Wire Transmitter
Versatile Mounting Styles
1 year warranty and 30 month pro-rated replacement
產品介紹
Exceptional Performance with the Differential Electrode Measurement Technique
This field-proven technique uses three electrodes instead of the two normally used in conventional pH sensors. Process and reference electrodes measure the pH differentially with respect to a third ground electrode. The end result is unsurpassed measurement accuracy, reduced reference junction potential, and elimination of sensor ground loops. These sensors provide greater reliability, resulting in less downtime and maintenance.
 
Lower Maintenance Needs with the Replaceable Salt Bridge
The salt bridge creates a barrier to contamination which minimizes the dilution of the internal standard cell solution. The result is lower maintenance needs and a longer time period between calibrations.
 
Built-in Preamp or Two-Wire Transmitter
The built-in preamp produces a strong signal, enabling you to locate the analyzer up to 3000 ft. (914 m) from the sensor. An optional built-in two-wire transmitter is available for applications requiring a 4-20 mA sensor signal. This option requires that the indicating instrument of the measuring system be capable of providing 24 VDC to power the sensor, and have adjustment means to calibrate for zero offset and span.
 
Versatile Mounting Styles
1.5 inch NPT threads are provided on both ends of the convertible mounting style sensor for either mounting into a pipe tee or attaching to the end of a pipe for immersion. The convertible style enables you to consolidate inventory, and thereby reduce associated costs. A union-mount style sensor and mounting tee are also available to conveniently install and remove the sensor for in-line service.
 
1 year warranty and 30 month pro-rated replacement
 
型號
 Conductors
 Body material
 Electrode Material
 Cable Length
 Mounting
6028P0 5-wire LCP Glass 3.0 m 10 ' 5-conductor (plus shield) Convertible (immersion or flow-through mount)
6028P033 5-wire LCP Glass 10 m 30 ' 5 conductor (plus shield) Convertible
6428P0 2-wire LCP Glass 3 m 10 ' 2-conductor (twisted pair) Convertible
6028P050 5-wire LCP Glass 50 ' Convertible
6028P020 5-wire LCP Glass 20 ' Convertible
6028P1 5-wire LCP Antimony 3 m 10 ' 2-conductor (twisted pair) Convertible
6058P0 5-wire LCP Glass 3 m 10 ' 2-conductor (twisted pair) Union mount
6022P0 5-wire Ryton Glass 3 m 10 ' 5 conductor (plus shield) Convertible
6028P025000010N 5-wire LCP Glass 25 ' Convertible
6028P015 5-wire LCP Glass 4.5 m 15 ' 5-conductor (plus shield) Convertible (immersion or flow-through mount)
6028P4 5-wire LCP Glass 3 m 10 ' Convertible (immersion or flow-through mount)
6028P210000010N 5-wire LCP Glass 10 ' Convertible
6022P020 5-wire Ryton Glass 6 m 20 ' 5 conductor (plus shield) Convertible
6028P090 5-wire LCP Glass 90 ' Convertible
6028P420 5-wire LCP Glass 20 ' Convertible
6028P510 5-wire LCP Glass 3 m 10 ' Convertible (immersion or flow-through mount)
6052P0 5-wire Ryton Glass 3 m 10 ' Union mount
6058P025 5-wire LCP Glass 25 ' Union
6028P012 5-wire LCP Glass 12 ' Convertible
60G9753-301          
6458P0 2-wire LCP Glass 3 m 10 ' 2-conductor (twisted pair) Union
6058P4 5-wire LCP Glass 10 ' Union
6028P01000A000N

 

5-wire LCP Glass 10 ' Convertible
6028P010F00000N

 

5-wire LCP Glass 10 ' Convertible
6028P010000010N

 

5-wire LCP Glass 10 ' Convertible
6028P6 5-wire LCP Glass 10 ' Convertible
6428P025 2-wire LCP Glass 25 ' Convertible
6058P1 5-wire LCP Antimony 10 ' Union
6422P0 2-wire Ryton Glass 3 m 10 ' 2-conductor (twisted pair) Convertible
6028P010000200N

 

5-wire LCP Glass 10 ' Convertible
6022P610 5-wire Ryton Glass 10 ' Convertible
6058P01000A000N

 

5-wire LCP Glass 10 ' Union
6022P050 5-wire Ryton Glass 50 ' Convertible
6022P1 5-wire Ryton Antimony 10 ' Convertible
6428P010010000N

 

2-wire LCP Glass 10 ' Convertible
6058P6 5-wire LCP Glass 10 ' Union
6428P020 2-wire LCP Glass 20 ' Convertible
6058P050 5-wire LCP Glass 50 ' Union
6058P610F00010N

 

5-wire LCP Glass 10 ' Union
6428P030 2-wire LCP Glass 30 ' Convertible
6428P050 2-wire LCP Glass 50 ' Convertible
6022P010000200N

 

5-wire Ryton Glass 10 ' Convertible
 
 
使用手冊 Type Language Size Date Edition
pH Differential Analog Sensors User Manual
DOC023.97.80076