The Intellical™ Sodium ISE Analysis Package contains all you need to start measuring Sodium (Na+) with your HQD meter instrument: The Intellical™ ISENA381 Sodium ISE with 1 m cable, the sodium ISAB powder pillows, 100 mg/L and 1000 mg/L sodium standard solutions.
Intellical™ 鈉 ISE 分析套組包含使用 HQD 測量鈉離子 (Na+) : 1 公尺長電纜的 Intellical™ ISENA381 鈉 ISE電極、ISAB 鈉測試粉枕包、100 mg/L 和 1000 mg/L 納標準品。
Parameter 分析參數 : Sodium
Method Type 方法 : Laboratory: Refillable Reference Element
Range 量測範圍 : 0.023 mg/L (1 x 10-6 M) - 23 g/L (1 M) Na⁺
Reference Type : Ag/AgCl
Accuracy : ±0.02 mV or 0.05%, whichever is greater
Resolution : Selectable up to 4 significant digits
Sensor Type : pH Glass
Temperature Range : 0 - 50 °C (32 - 122 °F)
Thermistor : ATC
Filling Solution : 0.02 M NH₄Cl (#2965126)
ISA Required : Sodium/Potassium ISA (#4451569)
關鍵字: 水 鈉離子 濃度 測試 電極, sodium probe, 水 鈉離子 分析, 鈉離子 濃度計
Intellical™ 鈉 ISE 分析套組包含使用 HQD 測量鈉離子 (Na+) : 1 公尺長電纜的 Intellical™ ISENA381 鈉 ISE電極、ISAB 鈉測試粉枕包、100 mg/L 和 1000 mg/L 納標準品。
Parameter 分析參數 : Sodium
Method Type 方法 : Laboratory: Refillable Reference Element
Range 量測範圍 : 0.023 mg/L (1 x 10-6 M) - 23 g/L (1 M) Na⁺
Reference Type : Ag/AgCl
Accuracy : ±0.02 mV or 0.05%, whichever is greater
Resolution : Selectable up to 4 significant digits
Sensor Type : pH Glass
Temperature Range : 0 - 50 °C (32 - 122 °F)
Thermistor : ATC
Filling Solution : 0.02 M NH₄Cl (#2965126)
ISA Required : Sodium/Potassium ISA (#4451569)
關鍵字: 水 鈉離子 濃度 測試 電極, sodium probe, 水 鈉離子 分析, 鈉離子 濃度計
Provides fast, stable, and accurate response
Refillable, easy to maintain design with a unique built-in-temperature sensor. Intellical™ digital probes lock on the result when the measurement is stable, removing the guesswork in having a moving number.
Ultimate traceability in measurement history
Stored time and date stamp for each measurement, operator and sample ID, calibration history, parameter, and probe serial number.
User alerted when re-calibration is needed
No guesswork needed to provide the most reliable and accurate results with each measurement.
Can be moved between meters without the need to re-calibrate or re-enter measurement settings
Ideal for a multi-user environment with multiple HQD series Laboratory and Portable meters.