水中導電度感測器 Contacting Conductivity Sensor, Mid-Range Conductivity (k=1.0), with ¾" PVDF Compression Fitting, 6 m
產品編號:3422C3A
Ultimate accuracy from ultra-pure to high conductivity applications.
從超純水到高導電度應用 - 極致精準監測
Applications 應用 : Boiler Water, Clean in Place (CIP), Membrane Protection
Contacting conductivity sensor with a nominal cell constant k = 1.0 cm-1, ¾-inch NPT PVDF compression fitting, 6 m analog cable, and titanium electrode.
Measuring range 量測範圍 : 0 - 2000 µS/cm.
Accuracy : ±2 % of reading above 200 µS/cm
Repeatability : ±0.5 % of reading
Response Time T90 : 30 s
Sensitivity : ±0.5 % of reading
Cell Constant k : 1 cm⁻¹
Flow : 0 - 3 m/s maximum, fully immersed
Temperature Measurement : -20 - 200 °C
Temperature Sensor : PT1000 RTD
Pressure Range : 1.7 bar at 150 °C
Transmission Distance : 100 m maximum;
1000 m maximum, when used with a termination box
關鍵字 : 純水 導電度分析儀, 監測 純水 導電度 濃度, conductivity analyzer, GLI conductivity sensor,
從超純水到高導電度應用 - 極致精準監測
Applications 應用 : Boiler Water, Clean in Place (CIP), Membrane Protection
Contacting conductivity sensor with a nominal cell constant k = 1.0 cm-1, ¾-inch NPT PVDF compression fitting, 6 m analog cable, and titanium electrode.
Measuring range 量測範圍 : 0 - 2000 µS/cm.
Accuracy : ±2 % of reading above 200 µS/cm
Repeatability : ±0.5 % of reading
Response Time T90 : 30 s
Sensitivity : ±0.5 % of reading
Cell Constant k : 1 cm⁻¹
Flow : 0 - 3 m/s maximum, fully immersed
Temperature Measurement : -20 - 200 °C
Temperature Sensor : PT1000 RTD
Pressure Range : 1.7 bar at 150 °C
Transmission Distance : 100 m maximum;
1000 m maximum, when used with a termination box
關鍵字 : 純水 導電度分析儀, 監測 純水 導電度 濃度, conductivity analyzer, GLI conductivity sensor,
High Performance Design
These sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a PT1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1 ºC accuracy.
These sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a PT1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1 ºC accuracy.
Resistivity and Conductivity Measurement Capability
These enhanced performance sensors measure from 0.057 µS/cm or 18.2 MΩ (theoretically pure water) up to 200,000 µS/cm.
These enhanced performance sensors measure from 0.057 µS/cm or 18.2 MΩ (theoretically pure water) up to 200,000 µS/cm.
Versatile mounting styles
Compression Fitting Sensors feature titanium electrodes and a compression fitting for universal installation with up to 102 mm insertion depth. The ½-inch or ¾-inch male NPT compression fitting are offered in PVDF or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor form the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 178 mm.
Compression Fitting Sensors feature titanium electrodes and a compression fitting for universal installation with up to 102 mm insertion depth. The ½-inch or ¾-inch male NPT compression fitting are offered in PVDF or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor form the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 178 mm.
Full-Featured “Plug and Play” with Digital Hach SC Controllers
There are no complicated wiring or set up procedures with any Hach SC controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”
There are no complicated wiring or set up procedures with any Hach SC controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”