水中導電度線上感測器 Digital Contacting Conductivity Sensor, Compression Fitting Style, Cell Constant 5.0
產品編號D3422D3
滿足超純水應用到高導電度應用的極精準監測需求 Ultimate accuracy from ultra-pure to high conductivity applications.

Nominal Cell Constant (K) = 5.0, 3/4 inch NPT PVDF Compression Fitting, 7m (23ft) Digital Cable, Titanium Electrode

Pressure Range : 0 - 300 psi (20.7 BAR)
Range : 0 - 10000 µS/cm
Repeatability : ± 0.5 of reading
Response Time : 90 % of reading within 30 seconds of step change
Sensitivity : ± .05 % of reading

關鍵字 : 純水 導電度 分析儀器, 純水 導電度 監測儀器
產品介紹
High Performance Design
These enhanced performance sensors are manufactured to exacting tolerances using high quality, rugged materials for demanding applications including ultra-pure water, clean-in-place (CIP), and boiler/condensate monitoring. Each sensor is tested to determine its unique, absolute four-digit cell constant. Also, each sensor has a Pt 1000 RTD temperature element built into its tip for exceptionally fast response to changes in temperature with ±0.1ºC accuracy.
 
Resistivity and Conductivity Measurement Capability
These enhanced performance sensors measure from theoretically pure water (0.057 µS/cm or 18.2 MΩ) up to 200,000 µS/cm. Hach’s sc Digital Controllers accept multiple digital sensor inputs, and can be user-set to measure conductivity, resistivity,TDS, salinity, or one of six calculated measurements.
 
Versatile Mounting Styles
Compression fitting sensors feature titanium electrodes and a compression fitting for universal installation with up to 4 inches (102 mm) insertion depth. The 1/2-inch or 3/4-inch male NPT compression fittings are offered in PVDF or 316 stainless steel. A longer version of this sensor is available for use with a 316 stainless steel ball valve hardware assembly to insert/retract the sensor from the process without stopping the flow. The longer version can also be used for insertion through a compression fitting. Maximum insertion depth is 7 inches (178 mm).
 
Full Featured “Plug and Play” with Digital SC Controller
There are no complicated wiring or set up procedures with any Hach SC controller. Just plug in any combination of Hach digital sensors and it's ready to use—it’s “plug and play.”